Technical specifications | ||
Enviroment | · Operating environment: · Temperature: 0~40℃ · Humidity:15%~80% · Pressure :86.0 kPa~106.0 kPa · Transport and Storage environment: · Temperature: -20~55℃ · Humidity:10%~90% · Pressure:86.0 kPa~106.0 kPa | |
Power Requirements | · AC: 100 ~240V.50Hz/60Hz · DC: Built-in rechargeable battery · Battery: 11.1V 24wh Li-ion Battery 2h operating after full charge 5min operating after low battery alarm | |
Application Range | • dult/Pediatric/Neonatal/Medicines/ Surgery • Operating Room/ICU/CCU | |
Date Storage | · Trend diagram/table: 720h · NIBP review:10000 events · Wave review: 12h · Alarm review:200 alarm events · Suppert drug concentration titration analysis | |
ECG | ·5 Leads: RA, LA, LL RL, V ·Lead mode: I,Ⅱ, Ⅲ,m., AVR, AVL, AVF.V ·Gain: x 1.25 mm/mV, 2.5 mm/mV. 5.0 mm/mV, 10 mm/mV, 20mm/mV (error ≤ 5%) | |
ST Segment | ·ST Segment Range: ﹣2.0 mV ~ + 2.0 mV ·Accuracy:0.02 mV | |
ECG. Sp02. NIBP. RESP. TEMP. PR.HR. | ||
Alarm | - Three Level: Low, medium and high - Indication: Auditory and visual - Setup: Default and custom - Silence: All alarms can be silenced - Volume: 45-85 dB measured at 1 meter | |
Thermal Recorder (Option) | - Method: Thermal dot array - Paper width:50 mm(1.97in) - Paper Speed:12.5/25/50(mm/sec) - Traces Maximum: 3 tracks | |
IBP (option) | - Channel: 2 - Measurement Range: -50~+300mmHg - Unit: mmHg, kPa - Accuracy:±2mmHgor2%,whichever is greater | |
EtC02(option, Microstream) | Range 0-19.7%(0-150 mmHg) - Unit: %, mmHg: kP - Respiration Rate Range 2-150 bpm | |
Optional accessories | · IBP · CO2 · Nellcor SpO2 | |
Standard accessories | · NIBP cuff & tube · ECG cable & electrodes · SpO2 sensor · TEMP probe · Lithium-ion battery · Power cable · Operator's manual |
Contact Person : | Maggie |
---|